Chart Industries Chart HALT/HASS Chambers HALT/HASS

 
Operation Tips

The following operational tips are provided to help suggest methods of making chamber performance more economical, or beneficial to our customers.

The following ideas will slightly alter chamber performance, but should help reduce nitrogen consumption:
When conducting a thermal ramp down, turn the heating event off.
When conducting a thermal soak at low temperatures, turn the heating event off.
When conducting a thermal ramp up, turn the cooling event off.
When conducting a long thermal soak at high temperatures with live device under test, experiment with leaving the heat event off.
Use multiple PID parameter sets to improve chamber performance based on products being tested.
To help reduce condensation on device under test, start the test profile with a short period of time doing a dry air or gaseous nitrogen purge of the chamber.
Adjust controller's high and low temperature safety limits for each test to reduce possibility of permanently damaging device under test.
Calibrate the controller and other related instrumentation regularly. This service can be done by Chart by calling our customer service group toll free 1-888-877-3093.
Trouble-Shooting
PROBLEM POSSIBLE CAUSE
Chamber controller doesn't power up Main disconnect not on.
Control power switch not on.
Ground fault interrupt tripped.
Controller circuit breaker tripped.
Doors won't open/close
(applies to Real-36 and Real-48)
Doors not closed far enough prior to turning switch.
Palm button pushed in.
Control power not on.
No air pressure to chamber.
Safety Mats Violated (If so equipped).
Fans don't come on Doors not closed.
Control power not on.
Fan Event not on.
Proximity switch(s) not engaged.
Chamber cools but doesn't heat High temp alarm activated.
Heat Event not on.
Controller internal limit exceeded.
Chamber heats but doesn't cool Low temp alarm activated.
Cooling Event not turned on.
Controller internal limit exceeded.
Chamber doesn't heat or cool Control power not on.
Fans not on.
Alarm condition exists.
Heating & Cooling events not on.
No vibration in manual mode Vibration Mode switch not in 'Manual' position.
Vibration Event not on.
No air supply to chamber.
Individual hammer air valves turned off.
Manual vibration set at zero.
Run button not set.

For further technical assistance, consult the supplied chamber manual, or call our technical service team at (888) 877-3093.